EasyManua.ls Logo

Anritsu MS2840A - :STATus:OPERation:ENABle <integer>; :STATus:OPERation:ENABle?

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.20 OPERation Status Register
2-981
2
SCPI Device Message Details
:STATus:OPERation:ENABle <integer>
Operation Status Register Enable
Function
This command sets the event enable register of the OPERation status
register.
Command
:STATus:OPERation:ENABle <integer>
Parameter
<integer>
Total byte of event enable register
Resolution 1
Range 0 to 65535
Details
This command is available only in the SCPI mode.
Example of Use
To set the event enable register of the OPERation status register to 16.
STAT:OPER:ENAB 16
:STATus:OPERation:ENABle?
Operation Status Register Enable Query
Function
This command queries the event enable register of the OPERation status
register.
Query
:STATus:OPERation:ENABle?
Response
<integer>
Parameter
<integer>
Total byte of event enable register
Resolution 1
Range 0 to 65535

Table of Contents

Other manuals for Anritsu MS2840A

Related product manuals