EasyManua.ls Logo

Anritsu MS2840A - [:SENSe]:SEMask:ATTenuation?

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.13 Spectrum Emission Mask
2-595
2
SCPI Device Message Details
[:SENSe]:SEMask:ATTenuation?
Spectrum Emission Mask Attenuator Query
Function
This command queries the attenuation when measuring the reference
power for Spectrum Emission Mask measurement.
Command
[:SENSe]:SEMask:ATTenuation?
Response
<rel_ampl>
Parameter
<rel_ampl>
Attenuation when measuring the reference
power
Range 0 to 60 dB
Resolution
[MS269xA]
2 dB
[MS2830A]
,
[MS2840A]
2 dB or 10 dB
Refer to “Table 2.4.2-3 Resolution of Input Attenuator” in the
MS2830A/MS2840A Signal Analyzer Operation Manual
(Spectrum Analyzer Function Operation).
Suffix code None. Value is returned in dB units.
Details
The setting range of this function is limited depending on the settings of
other parameters. Refer to the
MS2690A/MS2691A/MS2692A Signal
Analyzer Operation Manual (Spectrum Analyzer Function Operation)
or
MS2830A/MS2840A Signal Analyzer Operation Manual (Spectrum
Analyzer Function Operation)
for details.
Example of Use
To query the attenuation when measuring the reference power.
SEM:ATT?
> 10.00

Table of Contents

Other manuals for Anritsu MS2840A

Related product manuals