Gate View Frequency Mode
Function
This command selects the center frequency setting mode for Gate View.
Command
[:SENSe]:SWEep:EGATe:VIEW:FREQuency:AUTO <switch>
Parameter
<switch>
Center frequency setting mode
0|OFF
Sets the center frequency manually.
1|ON
Automatically sets the center frequency to that
for the frequency domain. (Default)
Details
This command is not available when Gate View is set to Off.
Example of Use
To manually set the center frequency for Gate View.
SWE:EGAT:VIEW:FREQ:AUTO OFF
[:SENSe]:SWEep:EGATe:VIEW:FREQuency:AUTO?
Gate View Frequency Mode Query
Function
To query the center frequency setting mode for Gate View.
Query
[:SENSe]:SWEep:EGATe:VIEW:FREQuency:AUTO?
Response
<switch>
Parameter
<switch>
Center frequency setting mode
0
The center frequency is set manually.
1
The center frequency for the frequency domain
is set automatically.
Example of Use
To query the center frequency setting mode for Gate View.
SWE:EGAT:VIEW:FREQ:AUTO?