Chapter 2 SCPI Device Message Detail
2-1002
[:SENSe]:MIXer:SIGNal ON|OFF|1|0
Signal Identifier
Function
This command enables/disables signal identifier operation, which
identifies measured signal and image signal when external mixer is used.
Command
[:SENSe]:MIXer:SIGNal <switch>
Parameter
<switch>
Signal ID
ON|1
Selects signal ID.
OFF|0
Does not select signal ID.
Default
OFF
Details
This function is available when Option 044/045 is installed for MS2830A.
This function is available when Option 044/046 is installed for MS2840A.
Signal ID cannot be set to On when the PS function is On.
Signal ID cannot be set to On when the Measure function is On.
Signal ID cannot be set to On when the Math function is other than Off.
Example of Use
To sweep while identifying measured signal and image signal.
MIX:SIGN ON