Chapter
Native Device Message Details
4-114
INIT:SPUR:PAUS:STAT?
Spurious Emission Pause Status Query
Function
This command queries whether the Spurious Emission measurement is
in remote control state and is being paused.
Query
INIT:SPUR:PAUS:STAT?
Response
switch
Parameter
switch
Status of the Spurious Emission measurement
1
Remote control state and paused
0
All other status
Details
It pauses before the pertinent segment is swept when Pause before
Sweep is set to On. This command is used to query if it is paused or not.
This command can only be used while it is in remote control state.
Pause is cancelled and measurement will stop when this command is
sent while it is in local status.
Example of Use
To query whether it is in remote control state and is being paused.
INIT:SPUR:PAUS:STAT?
> 0