SPUR:SWE:TIME:AUTO/SPUR:SWE:TIME:AUTO?
Spurious Emission Sweep Time Auto/Manual
Function
This command sets the sweep time of each segment to Auto/Manual.
Command
SPUR:SWE:TIME:AUTO
switch_1_com,switch_2_com,switch_3_com,switch_4_com,swit
ch_5_com,switch_6_com,switch_7_com,switch_8_com,switch_9
_com,switch_10_com,switch_11_com,switch_12_com,switch_13
_com,switch_14_com,switch_15_com,switch_16_com,switch_17
_com,switch_18_com,switch_19_com,switch_20_com
Query
SPUR:SWE:TIME:AUTO?
Response
switch_1_res,switch_2_res,switch_3_res,switch_4_res,swit
ch_5_res,switch_6_res,switch_7_res,switch_8_res,switch_9
_res,switch_10_res,switch_11_res,switch_12_res,switch_13
_res,switch_14_res,switch_15_res,switch_16_res,switch_17
_res,switch_18_res,switch_19_res,switch_20_res
Parameter
switch_n_com
Sweep time of Segment n Auto/Manual
ON|1
Sweep Time is set to Auto.
OFF|0
Sweep Time is set to Manual.
switch_n_res
Sweep time of Segment n Auto/Manual
1
Sweep Time is set to Auto.
0
Sweep Time is set to Manual.
Example of Use
To set the sweep time of each segment to Auto/Manual.
SPUR:SWE:TIME:AUTO
ON,ON,OFF,ON,ON,OFF,ON,ON,OFF,ON,ON,OFF,ON,ON,OFF,ON,ON,
OFF,ON,ON
SPUR:SWE:TIME:AUTO?
> 1,1,0,1,1,0,1,1,0,1,1,0,1,1,0,1,1,0,1,1