2.5 Signal Search
2-181
2
SCPI Device Message Details
:CALCulate:MARKer:PEAK:SORT:X
Search Peaks Sort X
Function
This command sorts as many markers as the number set in Search Peaks
Number by frequency (time) on the trace.
Command
:CALCulate:MARKer:PEAK:SORT:X
Details
This command is not available in the following cases:
During the Spurious Emission measurement AND when Displayed
Segment Mode is set to Auto.
During the Spectrum Emission Mask measurement.
When Blank is set for the active trace.
Example of Use
To sort the markers by frequency.
CALC:MARK:PEAK:SORT:X
:CALCulate:MARKer:PEAK:SORT:COUNt <integer>
Search Peaks Number
Function
This command sets the number of searches when executing Search Peaks
Sort Y/X.
Command
:CALCulate:MARKer:PEAK:SORT:COUNt <integer>
Parameter
<integer>
Number of searches
Range 1 to 10
Resolution 1
Default 10
Details
This command is not available during Spurious Emission measurement
and when Displayed Segment Mode is Auto.
Not available during Spectrum Emission Mask measurement.
Example of Use
To set the search number to 6.
CALC:MARK:PEAK:SORT:COUN 6