2.13 Spectrum Emission Mask
2-655
2
SCPI Device Message Details
[:SENSe]:SEMask:OFFSet[1]:LIST:RLEVel:AUTO
<switch_1>[,<switch_2>[,<switch_3>[,<switch_4>[,<switch_5>[,<switch_6>[,<
switch_7>[,<switch_8>[,<switch_9>[,<switch_10>[,<switch_11>[,<switch_12>]
]]]]]]]]]]
Spectrum Emission Mask Offset Reference Level Auto/Manual
Function
This command enables/disables the automatic reference level setting
function for the offset during Spectrum Emission Mask measurement.
Command
[:SENSe]:SEMask:OFFSet[1]:LIST:RLEVel:AUTO <switch_n>
Parameter
<switch_n>
Automatic reference level setting function
On/Off for Offset-n
ON|1
Enables the automatic reference level setting
function.
OFF|0
Disables the automatic reference level setting
function.
Details
The same value as the reference level when measuring the reference
power is used in the automatic setting.
Example of Use
To enable the automatic reference level setting function for the offset.
SEM:OFFS:LIST:RLEV:AUTO
ON,ON,OFF,OFF,ON,ON,ON,ON,OFF,OFF,ON,ON