2.7 Sweep/Trigger/Gate
2-305
2
SCPI Device Message Details
[:SENSe]:SWEep:EGATe:VIEW:BANDwidth[:RESolution] <freq>
Gate View Resolution Bandwidth
Function
This command sets the resolution bandwidth (RBW) for Gate View.
This parameter and the one for setting the RBW for time axis
measurement reference the same value.
Command
[:SENSe]:SWEep:EGATe:VIEW:BANDwidth[:RESolution] <freq>
Parameter
<freq>
Resolution bandwidth (RBW)
[MS269xA]
Range/Resolution 30 Hz to 31.25 MHz
However, the settable value is 30 Hz to 3 MHz
(for 1 to 3 sequences), 50 kHz, 5 MHz, 10 MHz,
or 20 MHz, or 31.25 MHz.
[MS2830A], [MS2840A]
Range/Resolution 30 Hz to 31.25 MHz
However, the settable value is 30 Hz to 3 MHz
(for 1 to 3 sequences), 500 Hz, 50 kHz, 2 MHz, 5
MHz, 10 MHz, 20 MHz, or 31.25 MHz.
[Common]
Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ
Hz is used when omitted.
Default Value that is set when RBW is set to Auto
Details
The automatic resolution bandwidth setting function is disabled when
the resolution bandwidth (RBW) is changed.
This command is not available when Gate View is set to Off.
Example of Use
To set the RBW for Gate View to 3 kHz.
SWE:EGAT:VIEW:BAND 3KHZ