2
SCPI Device Message Details
:CALCulate:BANDwidth|:BWIDth[:RESolution]:AUTO ON|OFF|1|0
Resolution Bandwidth Auto/Manual
Function
This command sets the resolution bandwidth (RBW) automatically.
When Couple Time/Freq. Domain is set to Off, each parameter is
specified in accordance with the already-specified frequency axis or time
axis measurement.
Command
:CALCulate:BANDwidth|:BWIDth[:RESolution]:AUTO <switch>
Parameter
MS269xA, MS2830A, MS2840A RBW Mode Normal
<switch>
Automatic setting
0|OFF
Sets the automatic setting to Off.
1|ON
Sets the automatic setting to On (Default).
MS2830A, MS2840A RBW Mode CISPR
(Only with MS2830A-016/116, MS2840A-016/116)
<switch>
Automatic setting (for CISPR RBW)
0|OFF
Automatic setting is set to Off.
1|ON
Automatic setting is set to On (Default).
Details
MS269xA, MS2830A, MS2840A RBW Mode Normal
This command is not available during the Spectrum Emission Mask
measurement or the Spurious Emission measurement.
MS2830A, MS2840A RBW Mode CISPR
Not available when the Measure function is set to On.
Not available when the Gate View function is set to On.
Example of Use
To enable the automatic setting of RBW.
CALC:BAND:AUTO ON
Related command
This command has the same function as the following commands.
[:SENSe]:BANDwidth|:BWIDth[:RESolution]:AUTO
[:SENSe]:ACPower:BANDwidth[:RESolution]:AUTO
[:SENSe]:CHPower:BANDwidth[:RESolution]:AUTO
[:SENSe]:OBWidth:BANDwidth[:RESolution]:AUTO
:CALCulate:ACPower:BANDwidth[:RESolution]:AUTO
:CALCulate:CHPower:BANDwidth[:RESolution]:AUTO
:CALCulate:OBWidth:BANDwidth[:RESolution]:AUTO