EasyManua.ls Logo

Anritsu MS2840A - :CONFigure:SANalyzer; :CONFigure?

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.8 Measure Function
2-323
2
SCPI Device Message Details
:CONFigure:SANalyzer
Measurement Off
Function
This command disables currently running measure function. No
operation is made if no measure function is running.
Command
:CONFigure:SANalyzer
Example of Use
To disable the currently running measure function.
CONF:SAN
:CONFigure?
Current Measurement Query
Function
This command queries the name of the current measure function.
Query
:CONFigure?
Response
<mode>
Parameter
<mode>
Measure function
ACP
ACP measurement
BPOW
Burst Average Power measurement
OBW
OBW measurement
CHP
Channel Power measurement
SEM
Spectrum Emission Mask measurement
SPUR
Spurious Emission measurement
SAN
OFF
Example of Use
To query the current Measure function.
CONF?
> ACP

Table of Contents

Other manuals for Anritsu MS2840A

Related product manuals