EasyManua.ls Logo

Anritsu MS2840A - [:SENSe]:SWEep:EGATe:VIEW:POINts?

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 SCPI Device Message Details
2-312
[:SENSe]:SWEep:EGATe:VIEW:POINts?
Gate View Trace Point Query
Function
This command queries the number of trace display points for Gate View.
This parameter and the one for setting the number of trace display
points for time axis measurement reference the same value.
Query
[:SENSe]:SWEep:EGATe:VIEW:POINts?
Response
<integer>
Parameter
<integer>
Number of trace display points
11
11 points
21
21 points
41
41 points
51
51 points
101
101 points
201
201 points
251
251 points
401
401 points
501
501 points
1001
1001 points
2001
2001 points
5001
5001 points
10001
10001 points
30001
30001 points (MS269xA Only)
Example of Use
To query the number of trace display points.
SWE:EGAT:VIEW:POIN?
> 2001