EasyManua.ls Logo

Anritsu MS2840A - SWE:EGAT:VIEW:FREQ:AUTO;SWE:EGAT:VIEW:FREQ:AUTO?

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter
Native Device Message Details
4-330
SWE:EGAT:VIEW:FREQ:AUTO/SWE:EGAT:VIEW:FREQ:AUTO?
Gate View Frequency Mode
Function
This command selects the center frequency setting mode for Gate View.
Command
SWE:EGAT:VIEW:FREQ:AUTO switch_com
Query
SWE:EGAT:VIEW:FREQ:AUTO?
Response
switch_res
Parameter
switch_com
Center frequency setting mode
OFF|0
Sets the center frequency manually.
ON|1
Automatically sets the center frequency to that
for the frequency domain.
switch_res
Center frequency setting mode
0
Sets the center frequency manually.
1
Automatically sets the center frequency to that
for the frequency domain.
Details
This command is not available when Gate View is set to Off.
Example of Use
To manually set the center frequency for Gate View.
SWE:EGAT:VIEW:FREQ:AUTO OFF
SWE:EGAT:VIEW:FREQ:AUTO?
> 0

Table of Contents

Other manuals for Anritsu MS2840A

Related product manuals