Chapter
4
Native Device Message Details
4-330
SWE:EGAT:VIEW:FREQ:AUTO/SWE:EGAT:VIEW:FREQ:AUTO?
Gate View Frequency Mode
Function
This command selects the center frequency setting mode for Gate View.
Command
SWE:EGAT:VIEW:FREQ:AUTO switch_com
Query
SWE:EGAT:VIEW:FREQ:AUTO?
Response
switch_res
Parameter
switch_com
Center frequency setting mode
OFF|0
Sets the center frequency manually.
ON|1
Automatically sets the center frequency to that
for the frequency domain.
switch_res
Center frequency setting mode
0
Sets the center frequency manually.
1
Automatically sets the center frequency to that
for the frequency domain.
Details
This command is not available when Gate View is set to Off.
Example of Use
To manually set the center frequency for Gate View.
SWE:EGAT:VIEW:FREQ:AUTO OFF
SWE:EGAT:VIEW:FREQ:AUTO?
> 0