2
SCPI Device Message Details
:FETCh:ACP[n]?
Adjacent Channel Power Read Fetch
Function
This command outputs the measurement result for Adjacent Channel
Power measurement.
Query
:FETCh:ACP[n]?
Response
When the result mode is “A”:
When Offset Setup Mode is Normal
When ACP measurement (ACP Result Type is OFFSet)
Two values of ref_carrier_a is output when ACP Reference is BSIDes.
<ref_carrier_a>(,<ref_carrier_a>),
<lower_offset_1_rel>,<lower_offset_1_abs>,
<upper_offset_1_rel>,<upper_offset_1_abs>,
<lower_offset_2_rel>,<lower_offset_2_abs>,
<upper_offset_2_rel>,<upper_offset_2_abs>,
<lower_offset_3_rel>,<lower_offset_3_abs>,
<upper_offset_3_rel>,<upper_offset_3_abs>
(in n=1 or when omitted)
In ACP measurement (when ACP Result Type is CARRier)
<integration_abs>,<total_carrier_a>,<power_1>,
<power_2>,<power_3>,<power_4>,<power_5>,<power_6>,
<power_7>,<power_8>,<power_9>,<power_10>,
<power_11>,<power_12>
(n=1 or when omitted)
In ACP measurement (when ACP Result Type is ALL)
<integration_abs>,<total_carrier_a>,
<power_1>,<power_2>,<power_3>,<power_4>,
<power_5>,<power_6>,<power_7>,<power_8>,
<power_9>,<power_10>,<power_11>,<power_12>,
<ref_carrier_a>(,<ref_carrier_a>),
<lower_offset_1_rel>,<lower_offset_1_abs>,
<upper_offset_1_rel>,<upper_offset_1_abs>,
<lower_offset_2_rel>,<lower_offset_2_abs>,
<upper_offset_2_rel>,<upper_offset_2_abs>,
<lower_offset_3_rel>,<lower_offset_3_abs>,
<upper_offset_3_rel>,<upper_offset_3_abs>
(n=1 or when omitted)