Chapter 2 SCPI Device Message Detail
2-730
[:SENSe]:SPURious:FSTop ON|OFF|1|0
Fail Stop
Function
This command sets whether to stop the measurement when a “Fail”
segment has been found.
Command
[:SENSe]:SPURious:FStop <switch>
Parameter
<switch>
Fail Stop On/Off
ON|1
Sets Fail Stop to On.
OFF|0
Sets Fail Stop to Off.
Example of Use
To stop the measurement when a “Fail” segment has been found.
SPUR:FSTop ON
[:SENSe]:SPURious:FSTop?
Fail Stop Query
Function
This command queries the On/Off state of Fail Stop.
Query
[:SENSe]:SPURious:FSTop?
Response
<switch>
Parameter
<switch>
Fail Stop On/Off
1
On
0
Off
Example of Use
To query the Fail Stop setting.
SPUR:FSTop?
> 1