2
SCPI Device Message Details
:CALCulate:BANDwidth|:BWIDth[:RESolution]:MODE?
Resolution Bandwidth Normal/CISPR Query
Function
This command queries the Resolution Bandwidth (RBW) mode. When
Couple Time/Freq. Domain is set to Off, each parameter is specified in
accordance with the already-specified frequency axis or time axis
measurement.
Query
:CALCulate:BANDwidth|:BWIDth[:RESolution]:MODE?
Response
<mode>
Parameter
<mode>
Resolution Bandwidth (RBW)
NORM
RBW (in normal cases)
CISP
CISPR RBW
Details
This function is available when MS2830A-016/116, MS2840A-016/116 is
installed.
Not available when the Measure function is set to On.
Not available when the Gate View function is set to On.
Example of Use
To query the RBW mode.
CALC
:
BAND:MODE?
> NORM
Related command
This command has the same function as the following commands.
[:SENSe]:ACPower:BANDwidth[:RESolution]:MODE?
[:SENSe]:CHPower:BANDwidth[:RESolution]:MODE?
[:SENSe]:OBWidth:BANDwidth[:RESolution]:MODE?
[:SENSe]:BPOWer|:TXPower:BANDwidth[:RESolution]:MODE?
:CALCulate:BANDwidth|:BWIDth[:RESolution]:MODE?
:CALCulate:ACPower:BANDwidth[:RESolution]:MODE?
:CALCulate:CHPower:BANDwidth[:RESolution]:MODE?
:CALCulate:OBWidth:BANDwidth[:RESolution]:MODE?
:CALCulate:BPOWer|:TXPower:BANDwidth[:RESolution]:MODE?