2.14 Spurious Emission Measurement Function
2-757
2
SCPI Device Message Details
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME
<seconds_1>,<seconds_2>,<seconds_3>,<seconds_4>,<seconds_5>,<seco
nds_6>,<seconds_7>,<seconds_8>,<seconds_9>,<seconds_10>,<seconds_
11>,<seconds_12>,<seconds_13>,<seconds_14>,<seconds_15>,<seconds_
16>,<seconds_17>,<seconds_18>,<seconds_19>,<seconds_20>
Spurious Emission Sweep Time
Function
This command sets the sweep time of each segment.
Command
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME <seconds_n>
Parameter
<seconds_n>
Sweep Time of Segment n
Range
[MS269xA]
2 ms to 1000 s
[MS2830A]
1 ms to 1000 s
[MS2840A]
1 ms to 1000 s
Suffix Code NS,US,MS,S
S is used when omitted.
Default Auto value
Example of Use
To set the sweep time of each segment.
SWE:TIME
0.1,0.1,0.1,0.1,0.1,0.1,0.1,0.1,0.1,0.1,0.2,0.3,0.1,0.1,
0.1,0.1,0.1,0.1,0.1,0.1