Chapter
4
Native Device Message Details
4-325
4
Native Device Message Details
SWE:EGAT:VIEW:BAND:AUTO/SWE:EGAT:VIEW:BAND:AUTO?
Gate View Resolution Bandwidth Auto/Manual
Function
This command automatically sets the resolution bandwidth (RBW) for
Gate View.
This parameter and the one for automatically setting the RBW for time
axis measurement reference the same value.
Command
SWE:EGAT:VIEW:BAND:AUTO switch_com
Query
SWE:EGAT:VIEW:BAND:AUTO?
Response
switch_res
Parameter
switch_com
Automatic setting
OFF|0
Disables the automatic setting.
ON|1
Enables the automatic setting.
switch_res
Automatic setting
0
Automatic setting is Off.
1
Automatic setting is On.
Details
This command is not available when Gate View is set to Off.
Example of Use
To enable the automatic setting of the RBW for Gate View.
SWE:EGAT:VIEW:BAND:AUTO ON
SWE:EGAT:VIEW:BAND:AUTO?
> 1