Chapter 2 SCPI Device Message Details
2-636
[:SENSe]:SEMask:DETector:OFFSet[:FUNCtion]
NORMal|POSitive|NEGative|SAMPle|RMS|AVERage
Spectrum Emission Mask Offset Detection Mode
Function
This command selects the detection mode of the waveform pattern of the
offset for Spectrum Emission Mask measurement in a lump sum.
Command
[:SENSe]:SEMask:DETector:OFFSet[:FUNCtion] <mode>
Parameter
<mode>
Detection mode selection for all offsets
NORMal
Simultaneous positive and negative peak
detection (Default)
POSitive
Positive peak detection
NEGative
Negative peak detection
SAMPle
Sample detection
RMS|AVERage
RMS detection
Example of Use
To set the detection mode of the offset.
SEM:DET:OFFS POS