Chapter 2 SCPI Device Message Details
2-166
:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10:MAXimum:POWer
Power Peak Search
Function
This command moves the active marker to the position where the peak
power of the zone width of the active marker becomes maximum in the
measurement band.
Command
:CALCulate:MARKer[n]:MAXimum:POWer
Parameter
<n>
Marker Number
1
Specifies marker 1
2
Specifies marker 2
3
Specifies marker 3
4
Specifies marker 4
5
Specifies marker 5
6
Specifies marker 6
7
Specifies marker 7
8
Specifies marker 8
9
Specifies marker 9
10
Specifies marker 10
When omitted: Specifies marker 1
Details
This cannot be executed under the following conditions:
● In Time Domain
● When Scale Mode is Linear
● When active trace is Blank
● When Spurious Emission measurement is On and Displayed Segment
Mode is Auto
● When Spectrum Emission Mask measurement is On
Example of Use
To move marker 1 to position where peak power of zone width is
maximum value.
CALC:MARK:MAX:POW