2
SCPI Device Message Details
:MMEMory:STORe:LLINe[1]|2|3|4|5|6 [<filename>[,<device>]]
Save Limit
Function
This command saves the Limit data settings to a CSV file.
Command
:MMEMory:STORe:LLINe[1]|2|3|4|5|6
[,<filename>[,<device>]]
Parameter
<filename> "
Name of file to save
"
String of up to 32 characters, not including the
extension, enclosed in double quotes (") or single
quotes (')
The following characters cannot be used:
\ / : * ? “ ” ‘ ’ < > |
When omitted it is saved as a csv file with the
following default file name: Limit
“Date”_“Sequential Number”.
<device>
Drive name
A,B,D,E,F,...
Defaults to D when omitted.
Details
• When the file name is omitted, a sequential number from 0 to 999 is
added to the default file name. When the sequential number reaches
999, no more files can be saved.
• Files are saved to the following directory in the specified drive.
\Anritsu Corporation\Signal Analyzer\User Data\Limit
• Up to 1000 files can be stored in the folder.
Example of Use
To save the Limit data "Limit" of Limit3 to E drive.
MMEM:STOR:LLIN3 "Limit",E