Chapter 2 SCPI Device Message Detail
2-1004
[:SENSe]:MIXer:SIGNal:MODE ISHift|ISUPpress
Signal Identifier Mode
Function
This command sets the type of signal identifier operation, which
identifies measured signal and image signal when external mixer is used.
Command
[:SENSe]:MIXer:SIGNal:MODE <type>
Parameter
<type>
Signal Identifier Mode
ISHift
Displays the sweep results of different polarities
alternately.
ISUPpress
Displays the sweep results of different polarities
after Minimum processing.
Default ISHift
Details
This function is available when Option 044/045 is installed for MS2830A.
This function is available when Option 044/046 is installed for MS2840A.
Example of Use
To set the type of signal identifier operation, which identifies measured
signal and image signal.
MIX:SIGN:MODE ISH