Chapter 2 SCPI Device Message Detail
2-744
[:SENSe]:SPURious:ATTenuation
<rel_ampl_1>|AUTO,<rel_ampl_2>|AUTO,<rel_ampl_3>|AUTO,<rel_ampl_4>
|AUTO,<rel_ampl_5>|AUTO,<rel_ampl_6>|AUTO,<rel_ampl_7>|AUTO,<rel_a
mpl_8>|AUTO,<rel_ampl_9>|AUTO,<rel_ampl_10>|AUTO,<rel_ampl_11>|AU
TO,<rel_ampl_12>|AUTO,<rel_ampl_13>|AUTO,<rel_ampl_14>|AUTO,<rel_a
mpl_15>|AUTO,<rel_ampl_16>|AUTO,<rel_ampl_17>|AUTO,<rel_ampl_18>|
AUTO,<rel_ampl_19>|AUTO,<rel_ampl_20>|AUTO
Spurious Emission Attenuator
Function
This command sets the attenuator value of each segment.
Command
[:SENSe]:SPURious:ATTenuation <rel_ampl_n>|AUTO
Parameter
<rel_ampl_n>
Attenuator of Segment n
Range 0 to 60 dB
Resolution
[MS269xA]
2 dB
[MS2830A]
,
[MS2840A]
2 dB or 10 dB
Refer to “Table 2.4.2-3 Resolution of Input Attenuator” in the
MS2830A/MS2840A Signal Analyzer Operation Manual
(Spectrum Analyzer Function Operation)
.
Suffix Code DB, dB is used even when omitted.
AUTO
Attenuator value is automatically set (Default).
Example of Use
To set the attenuator value.
SPUR:ATT
10DB,10DB,12DB,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,A
UTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO,AUTO