2.13 Spectrum Emission Mask
2-619
2
SCPI Device Message Details
[:SENSe]:SEMask:SWEep[1]:TIME?
Spectrum Emission Mask Sweep Time Query
Function
This command queries the sweep time when measuring the reference
power for Spectrum Emission Mask measurement.
Query
[:SENSe]:SEMask:SWEep[1]:TIME?
Response
<time>
Parameter
<time>
Sweep time
Range
[MS269xA]
2 ms to 1000 s
[MS2830A]
1 ms to 1000 s
[MS2840A]
1 ms to 1000 s
Resolution 1 ms (time ≤ 1 s)
0.1 s (1 s < time < 10 s)
1 s (10 s ≤ time)
Suffix code None. Value is returned in S units.
Details
The setting range of this function is limited depending on the settings of
other parameters. Refer to the
MS2690A/MS2691A/MS2692A Signal
Analyzer Operation Manual (Spectrum Analyzer Function Operation)
or
MS2830A/MS2840A Signal Analyzer Operation Manual (Spectrum
Analyzer Function Operation)
for details.
Example of Use
To query the sweep time when measuring the reference power.
SEM:SWE:TIME?
> 0.020000