Chapter 2 SCPI Device Message Details
2-208
:MMEMory:STORe:TRACe
TRACe1|TRACe2|TRACe3|TRACe4|TRACe5|TRACe6|GVIew|ALL[,<filenam
e>[,<device>]]
Save Wave Data
Function
This command saves the waveform data into a CSV file.
Command
:MMEMory:STORe:TRACe <trace>[,<filename>[,<device>]]
Parameter
<trace>
Trace to be saved
TRACe1
Trace A
TRACe2
Trace B
TRACe3
Trace C
TRACe4
Trace D
TRACe5
Trace E
TRACe6
Trace F
GVIew
Gate View
ALL
All traces.
<filename>
File name to be saved
String of up to 32 characters, not including the
extension, enclosed in double quotes (") or single
quotes (')
The following characters are not available.
\ / : * ? “ ” ‘ ’ < > |
When omitted, it is named “Wave Data
date_sequential number.csv”.
<device>
Drive name
A,B,D,E,F,...
Defaults to D when omitted.
Details
0 to 99 are added to a file when a file name is omitted. No more file can
be saved if 99 files are used.
Files are saved in the following directory of the specified drive.
\Anritsu Corporation\Signal Analyzer\User Data\TraceData\Spectrum
Analyzer
Up to 1000 files can be saved in the folder.
Trace A, B, C, D, E, and F cannot be set during the Spectrum Emission
Mask measurement.
Trace B, C, D, E, and F cannot be set during the Spurious Emission
measurement.