2.9 ACP
2-395
2
SCPI Device Message Details
[:SENSe]:ACPower:ADVanced:OFFSet:FILTer:TYPE
RECT|NYQuist|RNYQuist,RECT|NYQuist|RNYQuist,RECT|NYQuist|RNYQui
st,RECT|NYQuist|RNYQuist,RECT|NYQuist|RNYQuist,RECT|NYQuist|RNYQ
uist, RECT|NYQuist|RNYQuist,RECT|NYQuist|RNYQuist
Adjacent Channel Power Offset Filter Type
Function
This command sets offset filter for Adjacent Channel Power
measurement.
Command
[:SENSe]:ACPower:ADVanced:OFFSet:FILTer:TYPE
<mode_1>,<mode_2>,<mode_3>,<mode_4>,<mode_5>,<mode_6>,<m
ode_7>,<mode_8>
Parameter
<mode_n>
Filter type of Offset_n
RECT
Rectangular filter
NYQuist
Nyquist filter
RNYQuist
Root Nyquist filter (Default)
Details
This function can be set when the following trace is active.
Spectrum
To query the measurement results after executing this command, use the
*WAI command to control synchronization.
Example of Use
To set ACP measurement offset channel filter type to Nyquist and query
result.
ACP:ADV:OFFS:FILT:TYPE NYQ,NYQ,NYQ,NYQ,NYQ,NYQ,NYQ,NYQ
*WAI
FETC:ACP?