EasyManua.ls Logo

Anritsu MS2840A - :CALCulate:SGControl:POWer <numeric_value>

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 SCPI Device Message Detail
2-852
:CALCulate:SGControl:POWer <numeric_value>
Internal Signal Generator Control SG Output Level
Function
This command sets the SG output level when Internal Signal Generator
Control measurement is executed.
Command
:CALCulate:SGControl:POWer
Parameter
<numeric_value>
SG output level
Range 40.00 to +20.00 dBm (Start Frequency > 25 MHz) (*)
40.00 to +2.00 dBm (Start Frequency 25 MHz) (*)
136.00 to +15.00 dBm (Start Frequency > 25 MHz) (**)
136.00 to 3.00 dBm (Start Frequency 25 MHz) (**)
Resolution 0.01 dB
Default 40.00 dBm (*)
136.00 dBm (**)
Suffix code DBM, dBm is used even when omitted.
(*) without Option 022/122
(**) with Option 022/122
Details
This command is available only for MS2830A.
This function is available when MS2830A-052/152/352 is installed.
Example of Use
To set the SG output level to 30.00 dBm.
CALC:SGC:POW30.00DBM