Loading...
2.4 Marker
2-141
2
SCPI Device Message Details
:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10:PEAK:X?
Marker Frequency (Time) Query
Function
This command queries the frequency or time at the marker point. As for
the delta marker, this command queries the frequency difference or time
difference between the delta marker and the marker specified by
Relative To.
Query
:CALCulate:MARKer[n]:PEAK:X?
Response
<freq>
(in frequency domain)
<time>
(in time domain)
Parameter
<n>
Marker number
1
Marker 1
2
Marker 2
3
Marker 3
4
Marker 4
5
Marker 5
6
Marker 6
7
Marker 7
8
Marker 8
9
Marker 9
10
Marker 10
When omitted:
Marker 1
<freq>
In frequency domain
Resolution 0.01 Hz.
Suffix code None. Value is returned in Hz units.
999999999999 is returned when no
measurement is performed or an error occurs.
<time>
In time domain
resolution 0.1 ns.
Suffix code None. Value is returned in s units.
999999999999 is returned when no
measurement is performed or an error occurs.
Details
This command is not available during Spurious Emission measurement
and when Displayed Segment Mode is Auto.
Example of Use
To query the frequency at the marker point of Marker 1.
CALC:MARK:PEAK:X?
> 1.00