2.7 Sweep/Trigger/Gate
2-317
2
SCPI Device Message Details
[:SENSe]:SWEep:EGATe:HOLDoff?
Gate Hold Query
Function
This command queries the time during which gate input is disabled for a
set time, from the first gate signal input until the next gate signal input,
during gate sweep.
Query
[:SENSe]:SWEep:EGATe:HOLDoff?
Response
<time>
Parameter
<time>
Set time
Range 0 to 1 s
Resolution 10 ns
Suffix code None. Value is returned in s units.
Default 100 µs
Details
This command is not available for MS269x Series.
Example of Use
To query the gate input disable time.
SWE:EGAT:HOLD?
> 0.02000000
[:SENSe]:SWEep:EGATe:HOLDoff:STATe OFF|ON|0|1
Gate Hold On/Off
Function
This command switches On/Off the function to disable gate input for a
set time, from the first gate input signal until the next gate signal input,
during gate sweep.
Command
[:SENSe]:SWEep:EGATe:HOLDoff:STATe <switch>