Chapter
4
Native Device Message Details
4-259
4
Native Device Message Details
SEM:SWE:TIME:AUTO/SEM:SWE:TIME:AUTO?
Spectrum Emission Mask Sweep Time Auto/Manual
Function
This command sets the sweep time in measuring the reference power for
Spectrum Emission Mask measurement.
Command
SEM:SWE:TIME:AUTO switch_com
Query
SEM:SWE:TIME:AUTO?
Response
switch_res
Parameter
switch_com
Automatic setting of sweep time
ON|1
Enables the automatic setting function.
OFF|0
Disables the automatic setting function.
switch_res
Automatic setting of sweep time
1
Enables the automatic setting function.
0
Disables the automatic setting function.
Details
The setting range of this function is limited depending on the settings of
other parameters. Refer to
MS2690A/MS2691A/MS2692A Signal
Analyzer Operation Manual (Spectrum Analyzer Function Operation)
or
MS2830A/MS2840A Signal Analyzer Operation Manual (Spectrum
Analyzer Function Operation)
for details.
Example of Use
To set the sweep time in measuring the reference power automatically.
SEM:SWE:TIME:AUTO ON
SEM:SWE:TIME:AUTO?
> 1