2.13 Spectrum Emission Mask
2-705
2
SCPI Device Message Details
[:SENSe]:SEMask:RACouple OFF|ON|0|1
Couple Ref & ATT
Function
This command sets the shared setting for Reference Level and
Attenuator for Spectrum Emission Mask measurement to On/Off.
Command
[:SENSe]:SEMask:RACouple <switch>
Parameter
<switch>
Shared setting On/Off
ON|1
Sets the shared setting to On.(Default)
OFF|0
Sets the shared setting to Off.
Details
The values set in Reference Level and Attenuator of Reference Setup are
set to those of Offset Setup, when this function is set to On.
Example of Use
To set the shared setting of Reference Level and Attenuator for Spectrum
Emission Mask measurement to On.
SEM:RAC ON