Chapter
4
Native Device Message Details
4-205
4
Native Device Message Details
Details
This command is not available when Spurious Emission is set to On and
when Gate View is set to Off.
Example of Use
To set the reference level to 0 dBm.
RLV 0
RLV 0DBM
RLV?
> 0.00
RMK?
Reference Marker Position Query
Function
This command queries the marker position set by Relative To of the
active marker using displayed points from the screen left edge.
Query
RMK?
Response
position
Parameter
position
Marker position of the marker set by Relative To
(Number of displayed points from the screen left
edge)
Range 0 to 10, 20, 40, 50, 100, 200, 250, 400, 500,
1000, 2000, 5000, 10000, 30000 (MS269xA Only)
(The upper limit value varies according to the
number of trace display points.)
Resolution 1
Details
Returns *** when Marker Mode is set to other than Delta.
Returns the result in an absolute value even when Marker Mode of the
marker set by Relative To is set to Delta.
This command is not available during the Spurious Emission
measurement and when Displayed Segment Mode is set to Auto.
Example of Use
To query the marker position set by Relative To of the active marker.
RMK?
> 123