Chapter
4
Native Device Message Details
4-312
Details
This command is not available during the Spurious Emission
measurement and when Displayed Segment Mode is set to Auto.
Not available during Spectrum Emission Mask measurement.
Example of Use
To set the threshold value when detecting peak point to –10 dBm.
SRCHTHLVL -10
SRCHTHLVL?
> -10.00
ST/ST?
Sweep Time
Function
This command sets the sweep time. Each parameter is specified in
accordance with the already specified frequency axis or time axis
measurement.
Command
ST time
ST AUTO
Query
ST?
Response
time
No suffix code. Value is returned in
µ
s units.
Parameter
time
Sweep time
Range
[MS269xA]
2 ms to 1000 s
(For frequency axis measurement)
1
µ
s to 1000 s
(For time axis measurement)
[MS2830A], [MS2840A]
1 ms to 1000 s
(For frequency axis measurement)
1 ms to 999999 s
(For frequency axis measurement and when in
response)
1
µ
s to 1000 s (For time axis measurement)
Suffix code NS, US, MS, S
ms is used when omitted.