2.14 Spurious Emission Measurement Function
2-721
2
SCPI Device Message Details
[:SENSe]:SPURious:TYPE WORSt|EXAMine|PEAKs|FULL
Spurious Emission Result Type
Function
This command selects the measurement result type.
Command
[:SENSe]:SPURious:TYPE <type>
Parameter
<type>
Measurement result type
WORSt|EXAMine
Point with the least margin from the limit line is
displayed.
PEAKs|FULL
Spurious detected by the detection resolution
and the threshold set in each segment is
displayed.
Default Worst
Details
This command is not available during the Time Domain measurement.
Example of Use
To display the point with the least margin from the limit line.
SPUR:TYPE EXAM