Chapter 2 SCPI Device Message Detail
2-776
:CALCulate:SPURious[:RANGe][:LIST]:LIMit:ABSolute[:UPPer]:DATA:STOP
<ampl_1>,<ampl_2>,<ampl_3>,<ampl_4>,<ampl_5>,<ampl_6>,<ampl_7>,<a
mpl_8>,<ampl_9>,<ampl_10>,<ampl_11>,<ampl_12>,<ampl_13>,<ampl_14>
,<ampl_15>,<ampl_16>,<ampl_17>,<ampl_18>,<ampl_19>,<ampl_20>
Spurious Emission Limit Stop Level
Function
This command sets the absolute level limit of the stop frequency of each
segment for the Spurious Emission measurement.
Command
:CALCulate:SPURious[:RANGe][:LIST]:LIMit:ABSolute[:UPPer
]:DATA:STOP <ampl_n>
Parameter
<ampl_n>
Absolute level limit of stop frequency in Segment
n
Range –200 dBm to 50 dBm
Resolution 0.01 dB
Suffix Code DBM,DM
Default Auto value
Example of Use
To set the absolute level limit of stop frequency in each segment.
CALC:SPUR:LIM:ABS:DATA:STOP
−13,−13,−13,−13,−13,−13,-10,-10,−13,−13,−13,−13,−13,−13,
-10,-10,−13,−13,−13,−13