Chapter 2 SCPI Device Message Detail
2-870
:MEASure:BATCh:SEMask[n]? <filename>[,<device>]
Spectrum Emission Mask Batch Measure
Function
This command executes Spectrum Emission Mask measurement and
outputs the measurement result after the parameters described in the
specified parameter list file have been set.
Query
:MEASure:BATCh:SEMask[n]? <filename>[,<device>]
Response
The same value as for
:MEASure:SEMask[n]?
is returned.
Refer to
:MEASure:SEMask[n]?
for details.
Parameter
<filename>
Parameter list file
Specify with any character string enclosed by
double quotes (" ") or single quotes (' ').
<device>
Drive name
A,B,D,E,F,...
D drive is used when omitted.
Details
Spectrum Emission Mask measurement is executed and the result is
output after the parameters described in the specified parameter list
have been set.
The return values of this function vary depending on the result mode.
(cf.
:SYSTem:RESult:MODE
)
Place the parameter list files in the following folder on the specified
drive.
<device>:\Anritsu Corporation\Signal Analyzer\User Data\Batch
If the parameter list file has been changed, the changes must be reflected
with the
:MMEMory:RELoad:BATCh
command.
(cf.
:MMEMory:RELoad:BATCh
)