:TRIGger[:SEQuence]:FRAMe:PERiod?
Frame Trigger Period Query
Function
This command queries the frame trigger occurrence period.
Query
:TRIGger[:SEQuence]:FRAMe:PERiod?
Response
<time>
Parameter
<time>
Delay time until gate start
Range 1 µs to 1 s
Resolution 10 ns
Suffix code None. Value is returned in s units.
Details
This command is not available for MS269x Series.
This setting is shared by the trigger signal and gate signal. The same
value is also used when frame trigger is selected as the gate signal source
for gate sweep.
Example of Use
To query the frame trigger occurrence period.
TRIG:FRAM:PER?
> 0.02000000