Chapter
4
Native Device Message Details
4-162
MKL?
Marker Level Query
Function
This command queries the marker point level. In case of delta marker, it
queries the level ratio.
Query
MKL? marker
Response
level
Marker point level
No suffix code, in dB units, 0.001 dB
(When marker level display units are dB-system units)
No suffix code, in
µ
V units, 0.01 pV
(When marker level display units are V-system units)
No suffix code, in
µ
W units, 0.01 yW
(When marker level display units are W-system units)
No suffix code, in dB units, resolution 0.001 dB
(When Marker Mode is Delta and Scale Mode is Log)
No suffix code, No unit, range: 0.0000 to 10000, 0.0001 resolution
(When Marker Mode is Delta and when Scale Mode is Lin.)
Parameter
marker
Marker type
1
Marker1
2
Marker2
3
Marker3
4
Marker4
5
Marker5
6
Marker6
7
Marker7
8
Marker8
9
Marker9
10
Marker10
When omitted Active marker
Details
*** is returned when Marker Mode is Off.
This command is not available during the Spurious Emission
measurement and when Displayed Segment Mode is set to Auto.
Example of Use
To query the marker point level (1.234 dBm).
MKL?
> 1.234