2.7 Sweep/Trigger/Gate
2-285
2
SCPI Device Message Details
:TRIGger[:SEQuence]:HOLDoff?
Trigger Hold Query
Function
This command queries the time during which trigger input is disabled for
a set time from the first trigger input until the next trigger input.
Query
TRIGger[:SEQuence]:HOLDoff?
Response
<time>
Parameter
<time>
Offset time
Range 0 to 1 s
Resolution 10 ns
Suffix code None. Value is returned in s units.
Default 100 µs
Details
This command is not available for MS269x Series.
Example of Use
To query the gate input disable time.
TRIG:HOLD?
> 0.02000000
:TRIGger[:SEQuence]:HOLDoff:STATe OFF|ON|0|1
Trigger Hold On/Off
Function
This command sets on/off the function to disable trigger input for a set
time, from the first trigger input until the next trigger input.
Command
:TRIGger[:SEQuence]:HOLDoff:STATe <switch>
Parameter
<switch>
Trigger Hold On/Off
ON|1
Sets the Trigger Hold function to On.
OFF|0
Sets the Trigger Hold function to Off.