2.1 Frequency/Span
2-49
2
SCPI Device Message Details
[:SENSe]:FREQuency:BAND:MODE NORMal|SPURious
Frequency Band Mode
Function
This command sets the frequency band path. The frequency at which the
path is switched to the preselector band or a path that does not pass
through the preselector can be set with this function.
Command
[:SENSe]:FREQuency:BAND:MODE <mode>
Parameter
<mode>
Frequency band mode
[MS269xA]
NORMal
Sets the frequency to be switched to the
preselector band to 6.0 GHz.
SPURious
Sets the frequency to be switched to the
preselector band to 3.0 GHz.
[MS2830A-041/043/044/045]
NORMal
Sets the frequency to be switched to the
preselector band to 4.0 GHz.
SPURious
Sets the frequency to be switched to the
preselector band to 3.5 GHz.
[MS2840A-041/044/046]
NORMal
Sets the frequency to be switched to the
preselector band to 4.0 GHz.
SPURious
Sets the frequency to be switched to the
preselector band to 3.5 GHz.
Details
[MS269xA]
This command is not available when the Option 003 Preselector lower
limit frequency extension option is not installed.
[MS2830A]
This command is not available when used with the Option 040 3.6 GHz
Signal Analyzer.
[MS2840A]
This command is not available when used with the Option 040 3.6 GHz
Signal Analyzer.
Example of Use
To set the frequency to be switched to the preselector band to 6.0 GHz.
FREQ:BAND:MODE NORM