2.14 Spurious Emission Measurement Function
2-767
2
SCPI Device Message Details
[:SENSe]:SPURious:AVERage:COUNt:COUPle ON|OFF|1|0
Couple Storage Count
Function
This command sets the shared setting of the storage count in each
segment to On/Off.
Command
[:SENSe]:SPURious:AVERage:COUNt:COUPle <switch>
Parameter
<switch>
Shared setting On/Off
ON|1
Sets the Shared setting to On (Default).
OFF|0
Sets the Shared setting to Off.
Details
When set to On, the same as the storage count of the segment set in
Displayed segment is set to those of all other segments.
Example of Use
To set the shared setting of the storage count in each segment to On.
SPUR:AVER:COUN:COUP ON