2.17 Configuring Batch Measurement Settings
2-857
2
SCPI Device Message Details
:MMEMory:RELoad:BATCh [<device>]
Reloading Parameter List Files
Function
This command reflects the changes made to the parameter list files of the
specified drive.
Command
:MMEMory:RELoad:BATCh [<device>]
Parameter
<device>
Drive name
A,B,D,E,F,...
D drive is used when omitted.
Details
The parameter list files used for batch measurement are read in batch
when the main unit is started up and during application loading.
Therefore, even if the files are changed following startup (or following
loading), the changes are not applied to the measurement.(Measurement
is executed with the parameters before changes were made.)
This command has the effect of applying the changes made to the
parameter list files. The parameter list files as they exist at the moment
this command is sent are used for subsequent batch measurements.
Place the parameter list files in the following folder on the specified
drive.
<device>:\Anritsu Corporation\Signal Analyzer\User Data\Batch
This command supports updates of the parameter list files used for the
following commands. To update the parameter list file of another
application, first execute system change to that application, and then
send the update command.
:MEASure:BATCh:ACP[n]?
:MEASure:BATCh:OBWidth[n]?
:MEASure:BATCh:SEM[n]?
:MEASure:BATCh:SPURious[n]?
:MEASure:BATCh:IM?