2.18 Other Settings
2-959
2
SCPI Device Message Details
:CALibration:RCLock[:VALue] <integer>
Adjust Reference Clock
Function
This command sets the adjustment value for the internal reference signal
oscillator (Reference Clock).
Command
:CALibration:RCLock[:VALue] <integer>
Parameter
<integer>
Adjustment value
Range 0 to 1023 (MS2830A)
0 to 4095 (MS2840A)
Resolution 1
Example of Use
To set the adjustment value of the internal reference signal oscillator to
511
CAL:RCL 511
:CALibration:RCLock[:VALue]?
Adjust Reference Clock Query
Function
This command queries the adjustment value of the internal reference
signal oscillator (Reference Clock).
Query
:CALibration:RCLock[:VALue]?
Response
<integer>
Parameter
<integer>
Adjustment value
Range 0 to 1023 (MS2830A)
0 to 4095 (MS2840A)
Resolution 1
Example of Use
To query adjustment value 511 of the internal reference signal oscillator.
CAL:RCL?