Chapter 2 SCPI Device Message Details
2-318
Parameter
<switch>
Gate Hold On/Off
ON|1
Gate Hold is On
OFF|0
Gate Hold is Off
Details
This command is not available for MS269x Series.
The Gate Sweep (On/Off) function is automatically set to On when this
function is set to On.
Example of Use
To set the gate input disable time setting to On.
SWE:EGAT:HOLD:STAT ON
[:SENSe]:SWEep:EGATe:HOLDoff:STATe?
Gate Hold On/Off Query
Function
This command queries the on/off status of the function to disable trigger
input for a set time, from the first trigger input until the next trigger
input, during gate sweep.
Query
[:SENSe]:SWEep:EGATe:HOLDoff:STATe?
Response
<switch>
Parameter
<switch>
Gate Hold On/Off
1
Gate Hold is On
0
Gate Hold is Off
Details
This command is not available for MS269x Series.
Example of Use
To query the gate input disable time setting.
SWE:EGAT:HOLD:STAT?
> 1