Chapter
Native Device Message Details
4-83
4
Native Device Message Details
DPOINT/DPOINT?
Trace Point
Function
This command sets the number of trace display points. When Couple
Time/Frequency Domain is set to Off, each parameter is specified in
accordance with the already-specified frequency axis or time axis
measurement.
Command
DPOINT point
Query
DPOINT?
Response
point
Parameter
point
Number of trace display points
11
11 points
21
21 points
41
41 points
51
51 points
101
101 points
201
201 points
251
251 points
401
401 points
501
501 points
1001
1001 points
2001
2001 points
5001
5001 points
10001
10001 points
30001
30001 points (MS269xA Only)
Details
This command is not available during the Spectrum Emission Mask
measurement or the Spurious Emission measurement.
When Signal Generator Control is On, the number of trace display points
is set to 201 points.
Example of Use
To set the number of trace display points to 5001 points.
DPOINT 5001
DPOINT?
> 5001