EasyManua.ls Logo

Anritsu MS2840A - [:SENSe]:TOI:FREQuency:IM3:SEARch ON|OFF|1|0; :CALCulate:TOI:FREQuency:IM3:SEARch ON|OFF|1|0

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 SCPI Device Message Detail
2-842
[:SENSe]:TOI:FREQuency:IM3:SEARch ON|OFF|1|0
TOI 3rd Frequency Search On/Off
Function
This command turns on and off the search function for the third-order
distortion frequency in the TOI measurement.
Command
[:SENSe]:TOI:FREQuency:IM3:SEARch <switch>
Parameter
<switch>
Setting of the search function for the third-order
distortion frequency in the TOI measurement
ON|1
Turns on the search function for the third-order
distortion frequency.
OFF|0
Turns off the search function for the third-order
distortion frequency. (Default)
Example of Use
To turn on the search function for the third-order distortion frequency.
TOI:FREQ:IM3:SEAR 1
Related Command
This command has the same function as the following command.
:CALCulate:TOI:FREQuency:IM3:SEARch
:CALCulate:TOI:FREQuency:IM3:SEARch ON|OFF|1|0
TOI 3rd Frequency Search On/Off
Function
This command turns on and off the search function for the third-order
distortion frequency in the TOI measurement.
Refer to
[:SENSe]:TOI:FREQuency:IM3:SEARch.
Related Command
This command has the same function as the following command.
[:SENSe]:TOI:FREQuency:IM3:SEARch

Table of Contents

Other manuals for Anritsu MS2840A