2.18 Other Settings
2-957
2
SCPI Device Message Details
[:SENSe]:POWer[:RF]:PADJust <freq>
Pre-selector Tune
Function
This command sets the Preselector peaking bias value.
Command
[:SENSe]:POWer[:RF]:PADJust <freq>
Parameter
<freq>
Peaking bias value
Range –128 to 127
Resolution 1
Default 0
Details
[MS269xA]
Automatic setting of the peaking bias value cannot be used under the
following conditions:
•
When using the MS2690A:
•
When not using the preselector band:
The frequency which switched to the preselector band can be set in
Frequency Band Mode.
[MS2830A],
[MS2840A]
Automatic setting of the peaking bias value cannot be used under the
following conditions:
•
In case of MS2830A-040/041/043, MS2840A-040/041:
•
When not using the preselector band:
The frequency which switched to the preselector band can be set in
Frequency Band Mode.
Example of Use
To set the peaking bias value to 100.
POW:PADJ 100