2.13 Spectrum Emission Mask
2-649
2
SCPI Device Message Details
[:SENSe]:SEMask:OFFSet[1]:LIST:BANDwidth[1]:INTegration
<bandwidth_1>[,<bandwidth_2>[,<bandwidth_3>[,<bandwidth_4>[,<bandwidt
h_5>[,<bandwidth_6>[,<bandwidth_7>[,<bandwidth_8>[,<bandwidth_9>[,<ba
ndwidth_10>[,<bandwidth_11>[,<bandwidth_12>]]]]]]]]]]]
Spectrum Emission Mask Offset Integrate Bandwidth
Function
This command sets Integrate BW of the offset for Spectrum Emission
Mask measurement.
Command
[:SENSe]:SEMask:OFFSet[1]:LIST:BANDwidth[1]:INTegration
<bandwidth_n>
Parameter
<bandwidth_n>
Integrate BW of Offset-n
Range 30 Hz to 20 MHz
Resolution Resolutions are set as below, according to the
setting values of Integrate BW.
1 Hz (30 Hz to 1 kHz)
10 Hz (1 to 10 kHz)
100 Hz (10 to 100 kHz)
1 kHz (100 kHz to 1 MHz)
10 kHz (1 to 20 MHz)
Suffix code HZ,KHZ,KZ,MHZ,MZ,GHZ,GZ
Hz is used when omitted.
Details
The setting range of this function is limited depending on the setting of
RBW of the target offset. Refer to the
MS2690A/MS2691A/MS2692A
Signal Analyzer Operation Manual (Spectrum Analyzer Function
Operation)
or
MS2830A/MS2840A Signal Analyzer Operation Manual
(Spectrum Analyzer Function Operation)
for details.
This command is not available if detection of the target offset is set to
any of the following:
Pos&Neg
Positive
Negative
Sample
Example of Use
To set Integrate BW of the offset.
SEM:OFFS:LIST:BAND:INT
3MHZ,3MHZ,3MHZ,3MHZ,3MHZ,3MHZ,3MHZ,3MHZ,3MHZ,3MHZ,3MHZ,3
MHZ