EasyManua.ls Logo

Anritsu MS2840A - :INITiate:SPURious:PAUSe:STATe?

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 SCPI Device Message Detail
2-816
:INITiate:SPURious:PAUSe:STATe?
Spurious Emission Pause Status Query
Function
This command queries whether the Spurious Emission measurement is
in remote control state and paused or not.
Query
:INITiate:SPURious:PAUSe:STATe?
Response
<switch>
Parameter
<switch>
Status of the Spurious Emission measurement
1
Remote control state and paused
0
All other status
Details
It pauses before the pertinent segment is swept when Pause before
Sweep is set to On. This command is used to query if it is paused or not.
This command can only be used while it is in remote control state.
Pause is cancelled and measurement will stop when this command is
sent while it is in local status.
Example of Use
To query whether it is in remote control state and paused or not.
INIT:SPUR:PAUS:STAT?
> 0

Table of Contents

Other manuals for Anritsu MS2840A

Related product manuals