Native Device Message Details
4-232
SEM:OFFS:LIST:BAND:INT:AUTO/SEM:OFFS:LIST:BAND:INT:AUTO?
Spectrum Emission Mask Offset Integrate Bandwidth Auto/Manual
Function
This command sets Integrate BW of the offset for Spectrum Emission
Mask measurement automatically.
Command
SEM:OFFS:LIST:BAND:INT:AUTO
switch_1_com[,switch_2_com[,switch_3_com[,switch_4_com[,
switch_5_com[,switch_6_com[,switch_7_com[,switch_8_com[,
switch_9_com[,switch_10_com[,switch_11_com[,switch_12_co
m]]]]]]]]]]]
Query
SEM:OFFS:LIST:BAND:INT:AUTO?
Response
switch_1_res,switch_2_res,switch_3_res,switch_4_res,swit
ch_5_res,switch_6_res,switch_7_res,switch_8_res,switch_9
_res,switch_10_res,switch_11_res,switch_12_res
Parameter
switch_n_com
Automatic setting of Integrate BW of Offset-n
ON|1
Enables the automatic setting function.
OFF|0
Disables the automatic setting function.
switch_n_res
Integrate BW of Offset-n
1
Enables the automatic setting function.
0
Disables the automatic setting function.
Details
The setting range of this function is limited depending on the setting of
RBW of the target offset. Refer to
MS2690A/MS2691A/MS2692A Signal
Analyzer Operation Manual(Spectrum Analyzer Function Operation)
or
MS2830A/MS2840A Signal Analyzer Operation Manual (Spectrum
Analyzer Function Operation)
for details.
This command is not available when Detection of the target offset is set
to the following:
•
Pos&Neg
•
Positive
•
Negative
•
Sample